Specifications for IsoMed Constant Flow Infusion System
NOTE: Medtronic no longer sells the IsoMed constant flow infusion system. The information below is intended to help you manage patients who have this device.
| Radiopaque identifier |
ID Tag |
ID Tag |
ID Tag |
| Thickness |
24 mm |
29 mm |
37 mm |
| Weight (empty) |
113 g |
116 g |
120 g |
| Displacement volume |
73 mL |
95 mL |
133 mL |
| Diameter |
77 mm |
77 mm |
77 mm |
| Material |
Titanium |
Titanium |
Titanium |
| Capacity |
20 mL |
35 mL |
60 mL |
| Residual volume |
3.5 mL |
3.5 mL |
3.5 mL |
| Minimal flow rate |
.05/1.0/1.5 mL/ day |
.05/1.0/1.5 mL/ day |
.05/1.0/1.5 mL /day |
| Pump internal fluid path volume |
0.3 mL |
0.3 mL |
0.3 mL |
| Silicone septum puncture life |
500 |
500 |
500 |
| Bacterial retentive filter |
0.22 µ |
0.22 µ |
0.22 µ |
| Prime volume |
0.15 mL |
0.15 mL |
0.15 mL |
Indications
- Chronic intrathecal infusion of sterile, preservative-free morphine sulfate for chronic, intractable pain
How IsoMed Works
The IsoMed pump delivers a constant flow of fluid as the drug in the pump reservior is forced through flow-restrictive tubing by pressurized gas.
Standard flow rates are 0.5,1.0 and 1.5/mL per day. All IsoMed pumps have a screened catheter access port to allow direct access to the cerebrospinal fluid.
Manuals & Technical Resources
Find more information about the IsoMed system and how to manage patients with this device.