It is important to read this manual in its entirety before conducting a Magnetic Resonance Imaging (MRI) examination on a patient with any implanted Medtronic Deep Brain Stimulation (DBS) System component. Contact Medtronic if you have any questions.

Due to the number and variability of parameters that affect MRI compatibility, the safety of patients or continued functioning of Medtronic DBS Systems exposed to MRI cannot be absolutely ensured. MRI systems generate powerful electromagnetic fields that can produce a number of interactions with implanted components of the Medtronic DBS neurostimulation system. Some of these interactions, especially heating, are potentially hazardous and can lead to serious injury or death. However, with appropriate control measures, particularly with respect to the selection of MRI parameters and RF coils, it is generally possible to safely perform an MRI head scan on a Medtronic DBS patient. In addition, Medtronic DBS System components can affect the MRI image, potentially impacting the diagnostic use of this modality. The following information describes the potential interactions and control measures that should be taken to minimize the risks from these interactions.

Contraindication
(Back to top)

Implantation of a Medtronic DBS System is contraindicated for patients who will be exposed to Magnetic Resonance Imaging (MRI) using a full body transmit radio-frequency (RF) coil, a receive-only head coil, or a head transmit coil that extends over the chest area. Performing MRI with this equipment can cause tissue lesions from component heating, especially at the lead electrodes, resulting in serious and permanent injury including coma, paralysis, or death.

Warnings
(Back to top)

Precautions
(Back to top)

Note: The MRI guidelines provided here may significantly extend the MRI examination time or prevent some types of MRI examinations from being conducted on Medtronic DBS patients.

General Information on MRI
(Back to top)

An MRI system produces three types of electromagnetic fields that may interact with implanted neurostimulation systems. All three of these fields are necessary to produce an MRI image. Each of these fields can also produce specific but different types of interactions with implanted neurostimulator systems.

These fields include:

MRI Interactions with Implanted Medtronic DBS Systems
(Back to top)

MRI/neurostimulation system interactions are various, and the risk to the patient can range from minimal to severe. These interactions include the following:

Heating. The MRI RF field induces voltages onto the lead system that can produce significant heating effects at the lead-electrode– tissue interface or at the location of any breaks in the neurostimulator lead system. Component heating from the MRI RF field is the most serious risk from MRI exposure. Failure to follow these MRI recommendations can result in thermal lesions possibly resulting in coma, paralysis, or death.

Magnetic field interactions. Magnetic field interactions such as force and torque effects are produced by the static magnetic field. Any magnetic material will be attracted to the static magnetic field of the MRI. The force and torque effects may produce movement of the neurostimulator that can be uncomfortable to the patient, open a recent incision, or both. Medtronic DBS System components are designed with minimal magnetic materials.

Induced stimulation. Gradient magnetic fields may induce voltages onto the lead system that may cause unintended stimulation. The voltage of the induced stimulation pulses is proportional to the time rate of change (dB/dt) of the gradient pulses, the effective loop area created by the neurostimulator lead system, and the location of the lead system with respect to the gradient coils of the MRI.

Effects on neurostimulator function. The static, gradient, and RF fields of the MRI may affect the neurostimulator operation and programming. The static magnetic field may cause the neurostimulator to turn on or off if the neurostimulator uses a magnetically controlled switch that allows the patient to control stimulation by the application of a handheld magnet. Additionally, the MRI RF, static, and gradient fields may temporarily affect or disable other functions, such as telemetry or stimulation pulses. Parameters will need to be reprogrammed if the MRI causes a POR (Power On 8 English 2010-04 Reset) of the neurostimulator. Programmed parameters are retained for Activa RC Model 37612 and Activa PC Model 37601 Neurostimulators.

Image artifacts and distortion. The neurostimulation system components, particularly the neurostimulator, can cause significant imaging artifacts and/or distortion of the MRI image, particularly if the neurostimulator components contain magnetic material. The neurostimulator can cause the MRI image to be completely blocked from view (ie, signal loss or signal “void”) or severely distorted within several inches of the neurostimulator.

MRI Procedure
(Back to top)

Scope

These MRI/neurostimulator exposure guidelines apply to Medtronic DBS Systems comprising combinations of the following components:

Supervision

A responsible individual such as an MRI radiologist or MRI physicist must assure these procedures are followed. If the MRI is operated by an MRI technician, it is strongly recommended the responsible individual verifies that the MRI recommendations are followed.

Preparation

Do the following prior to performing an MRI examination on a Medtronic DBS patient:

  1. Inform the patient of the risks of undergoing an MRI.
  2. Check if the patient has any other implants or conditions that would prohibit or contraindicate an MRI examination. Do not conduct an MRI examination if any are found.
  3. Verify that all proposed MRI examination parameters comply with the "MRI Operation Settings." If not, the parameters must be modified to meet these requirements. If this cannot be done, do not perform an MRI.
  4. If the patient has implanted leads but does not have an implanted neurostimulator, perform the following steps:
  1. Wrap the external portion of the leads/percutaneous extensions with insulating material.
  2. Keep the external portion of the leads/percutaneous extensions out of contact with the patient.
  3. Keep the external leads/percutaneous extensions straight, with no loops, and running down the center of the head coil.
  1. If the patient has an implanted neurostimulator, perform the following steps:
  1. Review the neurostimulator with a clinician programmer and print out a copy of the programmed parameters for reference.
  2. Test for possible open circuits by measuring impedance and battery current on all electrodes in unipolar mode (see Table 1). If an open circuit is suspected, obtain an X-ray to identify whether the open circuit is caused by a broken lead wire. If a broken lead wire is found, do not perform an MRI.
  3. Table 1. Measurement Values Indicating Possible Open Circuits

    Note: For all devices, any measurement with a value of <250 Ω typically indicates a possible short circuit.

    Warning: An MRI procedure should not be performed in a patient with a Medtronic DBS System that has a broken lead wire because higher than normal heating may occur at the break or the lead electrodes, which can cause thermal lesions. These lesions may result in coma, paralysis, or death.

  4. If the Medtronic DBS System is functioning properly and no open or short circuits are found, program the neurostimulator to the settings provided in Table 2.
  5. Table 2. Recommended Neurostimulator Settings (for all programs) for MRI

    Caution: The decision to turn off a patient’s implanted neurostimulator in order to perform medical diagnostic or therapeutic procedures should be carefully considered based on the patient’s underlying medical condition. Consultation with the appropriate medical professionals (prescribing and implanting clinicians) is recommended.

MRI Operation Settings
(Back to top)

Prior to the MRI examination, a responsible individual such as an MRI radiologist or MRI physicist must assure the examination will be conducted according to the following MRI requirements. If standard MRI pulse sequences will be used, they must meet these requirements. If they do not, the pulse parameters must be adjusted so that they comply with these requirements:

Warning:

In-vitro testing has shown that exposure of the Medtronic DBS System to MRI under conditions other than described in this guideline can induce excessive heating at the lead electrodes or at breaks in the lead to cause lesions. These lesions may result in coma, paralysis, or death.

Contraindication

Implantation of a Medtronic DBS System is contraindicated for patients who will be exposed to Magnetic Resonance Imaging (MRI) using a full body transmit radio-frequency (RF) coil, a receiveonly head coil, or a head transmit coil that extends over the chest area. Performing MRI with this equipment can cause tissue lesions from component heating, especially at the lead electrodes, resulting in serious and permanent injury including coma, paralysis, or< death.

Warnings:

  • Ensure the SAR value is the value for head SAR. Some MRI systems may only display SAR, whole body SAR, or local body SAR. Make sure the value being limited is for head SAR. Excessive heating may occur if the wrong SAR value is used.
  • If MRI parameters must be manually adjusted after the initial automatic MRI prescan, do not make any adjustments that will increase the SAR value. Some MRI machines may not automatically update the displayed SAR value if manual adjustments are made. This may lead to higher than expected temperature increases in the Medtronic DBS System, particularly at the lead electrodes.
  • Limit the gradient dB/dt field to 20 Tesla/second or less.

Note: The recommendations provided are based on in-vitro testing and should result in a safe MRI examination of a patient with an implanted Medtronic DBS System. However, due to the many variables that affect safety, Medtronic cannot absolutely ensure safety or that the neurostimulator will not be damaged. The user of this information assumes full responsibility for the consequences of conducting an MRI examination on a patient with an implanted Medtronic DBS System.

Prior to the MRI Examination

Prior to the scan examination, the responsible individual must verify the MRI examination parameters comply with these guidelines.

During the MRI Examination

Post MRI Examination Review